Detail Hanna Turbidity And Chlorine Meter – HI83414

The HI83414 is a multiparameter instrument that measures the most important parameters in drinking water: turbidity and chlorine. The instrument is based on a state-of-the-art optical system which provides accurate results by minimizing stray light and color interferences. Periodic calibration with the supplied standards compensates for any variations in intensity of the tungsten lamp. The colorimeter portion of the meter uses a 525 nm narrow band interference filter for maintaining the proper wavelength in the measurement of free and total chlorine. All measurements are performed with 25 mm round cuvettes composed of special optical glass to ensure maximum repeatability of turbidity and chlorine measurements.

Turbidity Specifications

  • Non-Ratio Mode Range : 0.00 to 9.99; 10.0 to 40.0 NTU; 0.0 to 99.9; 100 to 268 Nephelos; 0.00 to 9.80 EBC
  • Non-Ratio Mode Resolution : 0.01; 0.1 NTU; 0.1; 1 Nephelos; 0.01 EBC
  • Ratio Mode Range
    0.00 to 9.99; 10.0 to 99.9; 100 to 4000 NTU; 0.0 to 99.9; 100 to 26800 Nephelos; 0.00 to 9.99; 10.0 to 99.9; 100 to 980 EBC
  • Ratio Mode Resolution
    0.01; 0.1; 1 NTU; 0.1; 1 Nephelos; 0.01; 0.1, 1 EBC
  • Range Selection
    automatic
  • Accuracy
    ±2% of reading plus 0.02 NTU (0.15 Nephelos; 0.01 EBC); ±5% of reading above 1000 NTU (6700 Nephelos; 245 EBC)
  • Repeatability
    ±1% of reading or 0.02 NTU (0.15 Nephelos; 0.01 EBC) whichever is greater
  • Stray Light
    < 0.02 NTU (0.15 Nephelos; 0.01 EBC)
  • Light Detector
    silicon photocell
  • Method
    nephelometric method (90°) or ratio nephelometric method (90° & 180°), adaptation of the USEPA Method 180.1 and Standard Method 2130 B
  • Measuring Mode
    normal, average, continuous
  • Turbidity Standards
    < 0.1, 15, 100, 750 and 2000 NTU
  • Calibration
    two, three, four, or five-point calibration
  • Free and Total Chlorine Specifications
  • Free and Total Chlorine Range
    0.00 to 5.00 mg/L (ppm)
  • Free and Total Chlorine Resolution
    0.01 mg/L (ppm) from 0.00 to 3.50 mg/L (ppm); 0.10 above 3.50 mg/L (ppm)
  • Free and Total Chlorine Accuracy
    ±0.02 mg/L @ 1.00 mg/L
  • Light Detector
    silicon photocell with 525 nm narrow band interference filter
  • Method
    adaptation of the USEPA Method 330.5 and Standard Method 4500-Cl G.
  • Standards
    1.00 mg/L (ppm) free chlorine; 1.00 mg/L (ppm) total chlorine
  • Calibration
    one-point calibration

General Specifications

  • Light Source
    tungsten filament lamp / greater than 100,000 readings
  • Display
    40 x 70 mm graphic LCD (64 x 128 pixels) with backlight
  • Log Memory
    200 records
  • Connectivity
    USB
  • Environment
    0 to 50°C (32 to 122°F), RH max 95% non-condensing
  • Power Supply
    115/130 VAC; 50/60 Hz
  • Dimensions
    230 x 200 x 145 mm (9.0 x 7.9 x 5.7”)
  • Weight
    2.5 kg (88 oz.)