PT. SRIAYU MULTI KREASI

Jln. Hayam Wuruk No. 2-5

Gedung Harco Glodok Lantai 5 Blok A.42 - Jakarta Barat

PT. SRIAYU MULTI KREASI

Hanna HI96752 Calcium And Magnesium Portable Photometer


Detail Hanna HI96752 Calcium And Magnesium Portable Photometer

Details : 

The HI96752 portable photometer is for the measurement of calcium and magnesium. Hanna’s portable photometers feature an advanced optical system; the combination of a LED light source, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

Features at-a-glance

CAL Check™ – Allows for performance verification and calibration of the meter using NIST traceable standards.

GLP – Review of the last calibration date.

Auto-shut off – Automatic shut off after 10 minutes of non-use when the meter is in measurement mode. Prevents wastage of batteries in the event the meter is accidentally left on.

Battery status indicator – Indicates the amount of battery life left.

Built-in timer – Display of time remaining before a measurement is taken. Ensures that all readings are taken at the appropriate reaction intervals for the test being performed.

Error messages – Messages on display alerting to problems including no cap, high zero, and standard too low.

Cooling lamp indicator – To maintain the desirable wavelength to be used for absorbance, it is necessary to ensure components are not overheated from the heat generated by the tungsten lamp. Each photometer is designed to allow a minimal amount of time for components to cool. The cooling lamp indicator is displayed prior to a reading being taken.

Units of measure – Appropriate unit of measure is displayed along with reading.

Absorbance Specifications
Light Source
light emitting diode
Light Detector
silicon photocell with narrow band interference filter @ 466 nm

Additional Specifications
Calcium Range
0 to 400 mg/L (ppm)
Calcium Resolution
1 mg/L
Calcium Accuracy @ 25°C (77°F)
±10 mg/L ±5% of reading
Magnesium Range
0 to 150 mg/L (ppm)
Magnesium Resolution
1 mg/L
Magnesium Accuracy @ 25°C (77°F)
±3 mg/L ±3% of reading
Method
calcium: adaptation of oxalate method; magnesium: adaptation of the calmagite method
Auto-off
after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
Environment
0 to 50°C (32 to 122°F); RH max 95% non-condensing
Weight
360g (12.7oz.)
Battery Type
9V battery
Dimensions
193 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Ordering Information
HI96752 is supplied with sample cuvettes with caps (2), 9V battery, instrument quality certificate and instruction manual.
CAL Check™ standards and testing reagents sold separately.

Have any Question or Comment?