PT. SRIAYU MULTI KREASI

Jln. Hayam Wuruk No. 2-5

Gedung Harco Glodok Lantai 5 Blok A.42 - Jakarta Barat

PT. SRIAYU MULTI KREASI

Hanna ISO 7027 Compliant Benchtop Turbidity Meter


Detail Hanna ISO 7027 Compliant Benchtop Turbidity Meter

The HI88713 Precision ISO Turbidity Benchtop Meter is specially designed for water quality measurements, providing reliable and accurate readings, even within low turbidity ranges. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimizes stray light and color interferences. Periodic calibration with the supplied standards compensates for any variations in intensity of the tungsten lamp. The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements.

Turbidity Specifications

  • FNU Range
    0.00 to 9.99 FNU, 10.0 to 99.9 FNU, 100 to 1000 FNU
  • FNU Resolution
    0.01 FNU, 0.1 FNU, 1 FNU
  • FNU Accuracy
    ±2% of reading plus stray light
  • FAU Range
    10.0 to 99.9 FAU, 100 to 4000 FAU
  • FAU Resolution
    0.1 FAU, 1 FAU
  • FAU Accuracy
    ± 10% of reading
  • Non-Ratio Mode Range
    0.00 to 9.99 NTU; 10.0 to 99.9 NTU; 100 to 1000 NTU 0.00 to 9.99 EBC; 10.0 to 99.9 EBC; 100 to 245 EBC
  • Non-Ratio Mode Resolution
    0.01; 0.1; 1 NTU; 0.01; 0.1; 1 EBC
  • Non-Ratio Mode Accuracy
    ±2% of reading plus stray light
  • Ratio Mode Range
    0.00 to 9.99 NTU; 10.0 to 99.9 NTU; 100 to 4000 NTU 0.00 to 9.99 EBC; 10.0 to 99.9 EBC; 100 to 980 EBC
  • Ratio Mode Resolution
    0.01; 0.1; 1 NTU; 0.01; 0.1; 1 EBC
  • Ratio Mode Accuracy
    ±2% of reading plus stray light, ±5% of reading above 1000 NTU
  • Range Selection
    automatic
  • Repeatability
    ±1% of reading or 0.02 NTU (0.15 Nephelos; 0.01 EBC) whichever is greater
  • Stray Light
    < 0.02 NTU (0.15 Nephelos; 0.01 EBC)
  • Light Detector
    silicon photocell
  • Method
    nephelometric method (90°) or ratio nephelometric method (90° & 180°), adaptation of the USEPA Method 180.1 and Standard Method 2130 B
  • Measuring Mode
    normal, average, continuous
  • Turbidity Standards
    < 0.1, 15, 100, 750 and 2000 NTU
  • Calibration
    two, three, four, or five-point calibration
  • General Specifications
  • Light Source
    IR LED
  • Display
    40 x 70 mm graphic LCD (64 x 128 pixels) with backlight
  • Log Memory
    200 records
  • Connectivity
    USB
  • Environment
    0 to 50°C (32 to 122°F), RH max 95% non-condensing
  • Power Supply
    12 VDC adapter (included)
  • Dimensions
    230 x 200 x 145 mm (9.0 x 7.9 x 5.7”)
  • Weight
    2.5 kg (88 oz.)