Detail Hioki Memory Hilogger LR8400-20

  • Compact size despite 30-channel standard capabilities
  • Expand up to 30 additional channels
  • Protected againts unexpected power outages
  • Write data to USB memory stick or CF card in real time
  • Built-in with USB 2.0 or 100 BASE-TX LAN interfaces
  • 5.7 TFT color display

Specifications :

1. Analog input Universal unit LR8501 Actual functionality will depend on combination of units installed Note: Isolated from each channel to chassis :
[No. of channels] 15 analog channels isolated scanning method input (4 terminals: push-button type)
[Voltage] ±10 mV to ±100 V, 1-5 V f.s. Max. resolution: 500 nV, (Isolated between channels and from each channel to chassis)
[Temperature: Thermocouples] -200 ˚C to 2000 ˚C (depends on sensor), Thermocouples (K, J, E, T, N, R, S, B, W,) Max. resolution 0.01 ˚C, (Isolated between channels and from each channel to chassis)
[Temperature: Pt 100, JPt 100 sensor] -200 ˚C to 800 ˚C, Max. resolution 0.01 ˚C, (Not isolated between channels)
[Resistance] 0 Ω to 200 Ω f.s. Max. resolution 0.5 mΩ, (Not isolated between channels)
[Humidity] 5.0 to 95.0 % rh (use with optional sensor), resolution 0.1 % rh, (Not isolated between channels nor from each channel to chassis)
[Max. rated voltage between isolated input channels] 300 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC

2. Analog input Votage/temp unit LR8500 Actual functionality will depend on combination of units installed Note: Isolated from each channel to chassis
[No. of channels] 15 analog channels; isolated scanning method input (2 terminals: M3 screw type)
[Voltage] ±10 mV to ±100 V, 1-5 V f.s. Max. resolution: 500 nV, (Isolated between channels and from each channel to chassies)
[Temperature: Thermocouples] -200 ˚C to 2000 ˚C (depends on sensor), Thermocouples (K, J, E, T, N, R, S, B, W), Max. resolution 0.01 ˚C, (Isolated between channels and from each channel to chassies)
[Pt 100, JPt 100 sensor] [Resistance] Not available
[Humidity] 5.0 to 95.0 % rh (use with optional sensor), resolution 0.1 % rh, (Not isolated between channels nor from each channel to chassies)
[Max. rated voltage between isolated input channels] 250 V DC
[Max. allowable input] ±100 V DC
[Max. rated voltage from isolated terminals to ground] 300 V AC, DC
Caution: Max. voltage from terminals to ground without damage

3. Pulse, Digital input : [No. of channels] 8 channels, pulse / digital selectable for each channel, M3 screw terminal, not isolated, common ground
[Pulse totalization] 0 to 1000 M pulse, 1 range (No-voltage ‘a’ contact; normally open, open collector or voltage input), Max. resolution 1 pulse
[Rotation count] 0 to 5000 /n (r/s) f.s. 1 range (same as Pulse totalization input signal condition), resolution 1/n (r/s) Note: “n” is the number of sensor output pulses per revolution, 1 to 1000
[Digital input] Record logical “1” or “0” at each sampling
[Max. rated voltage between input channels] Not isolated
[Max. allowable input] 0 to 50 V

4. Recording intervals : 10 ms to 50 ms, 100 ms to 1 hour, 19 selections (All input channels are scanned within each recording interval.)
Note: limited by using channels at 10 ms to 50 ms interval

5. Digital filter
Select from OFF/ 50 Hz/ 60 Hz (the cut-off frequency is automatically set)

6. Data storage
Internal memory: 8 M-words, Data storage media: CF card or USB memory (Only data recorded to a genuine HIOKI CF card is guaranteed)

7. LAN interfaces
100BASE-TX, Functions: Data acquisition using bundled software or PC commands, FTP server, FTP client, HTTP server function, or E-mail system
Note: LAN communication support planned from software Ver. 1.20

8. USB interfaces
USB 2.0 High-speed capable, series mini-B receptacle
Functions: Data acquisition using bundled software or PC commands, Transfer data from the CF card to a PC via USB drive mode
Note: Data transfer not possible from USB memory sticks

9. Display device
5.7 inch TFT color liquid crystal display (640 × 480 pixel)

10. Other functions
Save waveform data in real time to the CF card or USB memory stick,
Numerical value calculations, Waveform calculations, and others

11. AC Power supply
Using the AC adapter 9418-15 (100 to 240 V AC, 50/60 Hz), 7 VA

12. DC Power supply
Using the Battery pack Z1000 (optional accessory), Continuous use 5 hr,
External power: 10 to 28 V DC (Please contact your HIOKI distributor for connection cord)

13. Dimensions and Mass
272 mm (10.71 in) W × 182.4 mm (7.18 in) H × 66.5 mm (2.62 in) D,
1.8 kg (63.5 oz), (LR8400-20 main unit, excluding the Battery Pack 370 g/ 13.1 oz)

14. Accessories
Instruction manual ×1, Measurement guide ×1, CD-R (data collection software “Logger Utility”) ×1, USB cable ×1, AC adapter 9418-15 ×1