Detail Hioki Safety Hitester 3258
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)Regarding DMM Accuracy Due to the many ranges and functions available in a DMM, only the basic accuracy is indicated for reference. Please refer to the individual catalogs for detailed accuracy information.
Measurement parameter
AC voltage (AC potential bridge)
Objects of measurement
Insulated conductors (IV or CV equivalent, min. 100 mm² x-section), metal conductors Note: Not usable on shielded conductors
Measurement range
420.0V/ 600V AC (Auto)
Measurement values
True RMS
Max. rated voltage to earth
600V AC rms
Influence of adjacent wiring
±5% rdg. or less
Temperature characteristics
0.05 % rdg./ °C
Display
Digital LCD, max. 4200 digit (zero suppress at 10 digits or less)
Display refresh rate
Approx. once every 0.6 seconds
Display response
2.4 s or less
Functions
Data hold function, auto-power off, low battery warning
Power supply
LR6 (AA) alkaline battery ×6, Continuous use: 14 hr (Power ON, no measurement), or 1-minute measurements for 1000 times, or 5-minute measurements for 200 times
Dimensions and mass
51mm (2.01in) W × 275mm (10.83in) H × 37.5mm (1.48in) D (one probe), 670g (23.6oz)
Accessories
Instruction manual ×1, LR6 (AA) alkaline battery ×6, Soft carrying case ×1