Detail HUKSEFLUX – MTN02
Product Description
The MTN02 performs fast measurements of the thermal conductivity and resistivity of soils. Measurement with MTN02 complies with ASTM D5334 and IEEE 442 standards. The system is primarily designed for laboratory (indoor) measurements but can also be used for on-site (field) measurements. MTN02 is operated and powered from the handheld control and readout unit CRU02.
Specification
Measurands : -thermal conductivity -thermal resistivity -temperature
Measurement range (λ) : 0.1 to 6 W/(m•K) (all common soils)
Rated operating temperature range : from 0 to +50 °C
Measurement method : absolute measurement per ASTM D 5334-08 and IEEE Standard 442-1981(03)
Data analysis : initial analysis by CRU, final review of measurement on PC (required by ASTM)
Uncertainty (at 20 °C) : +/- (6 % of reading + 0.04) W/(m•K)
Heating interval : 300 s (typical)
Data storage capacity : 50 measurements
Length TP07 : 0.12m
Diameter TP07 : 3.5 x 10-3 m
Data communication : USB